Press Mode key ( ) for the Maximum (MAX), Minimum (MIN), Difference between MAX and MIN (DIF)
and Average (AVG) modes. During the measurement, the modes reading will be displayed beside the
mode icon.
Simply aim the thermometer at the desired target
and press the trigger (Meas. key ) to display the
surface temperature. The Distance:Spot ratio is
11:1, which means the target diameter will be 1-inch
at an 11-inch distance from the target. Please make
sure the target area is within the field of view.
Made in Taiwan
T
T
N
N
4
4
0
0
8
8
L
L
T
T
T
T
h
h
e
e
r
r
m
m
o
o
m
m
e
e
t
t
e
e
r
r
O
O
p
p
e
e
r
r
a
a
t
t
i
i
n
n
g
g
I
I
n
n
s
s
t
t
r
r
u
u
c
c
t
t
i
i
o
o
n
n
s
s
The thermometer is a non-contact infrared thermometer. There are many mathematical modes for the Infrared function. Please
remember to keep away from children and don’t use it
for safety related applications.
(Default Screen)
Simply aim the thermometer at the measure target
with Lens (
7
) and press Meas. key (
9
) to display
the surface temperature.
The Distance:Spot is 11:1. Please make sure the
target area is within the field of view.
FUNCTION
Press Emissivity key (
1
) for setting the emissivity.
Press Emissivity key (
1
), then press Up key (
4
) or Down key (
2
) to set the emissivity, then press Mode ke
y
(
5
) to confirm it. The emissivity can be changed from 0.10 (10E) to 1 (100E).
Press Mode key (
5
) for scrolling more display function as follows.
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key (
5
) for the Maximum (MAX), Minimum (MIN), Different between MAX and MIN (DIF) and
Average (AVG) modes. Durin
g
the measurement, the special modes readin
g
will be displa
y
ed beside the
mode icon.
Press Up key (
4
) or Down key (
2
) key to change the High Alarm (HAL) or Lo Alarm (LAL), then press Meas.
key (
9
) to confirm it. For example: When the reading 27
o
C < LAL 27.1
o
C, the Low icon will flash and
y
ou will hear
a beep sound.
Connect the thermocouple with Thermocouple socket (
10
) and put the probe in/on the tar
g
et, the thermometer will
display the temperature automatically without pressin
g
an
y
button. To see the minimum or maximum data durin
g
the probe measurement, please hold down the Up key (
4
) or Down key (
2
).
After measure high temp, the probe may remain HOT for a while.
** The thermometer will automatically shut off if left idle for more than 60sec, unless in PRB mode. (In PRB mode, it will shut off if
left idle for more than 12 minutes.)
ADD VALUE
Press Up key (
4
) for LOCK mode ON/OFF. The lock mode is particularl
y
useful for continuous
monitoring of temperatures for up to 60 minutes.
In MAX, MIN, DIF, AVG
mode:
Press Down key (
2
) for
o
C or
o
F transferred.
and press Up key (
4
) for backlight function ON/OFF.
In all modes: First
hold on the Meas. key
(
9
)
and press Down key (
2
) for laser function ON/OFF.
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING The thermometer should be stored at room temperature between -4 to 9°F (-20 to 65°C).
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times. Care should be taken when
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol allowing the lens to fully dry before using the
thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lois displa
y
ed when the temperature bein
g
measured is outside of the settin
g
s of
HAL and LAL (high and low alarms).
‘Er2’ is displa
y
ed when the thermometer is exposed to rapid chan
g
es in the ambient
temperature. ‘Er3’ is displayed when the ambient temperature exceeds 32°F (0°C) or 122°F
(50°C). The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize
to the working/room temperature.
For all other error messa
g
es it is necessar
y
to reset the thermometer. To reset it, turn the
instrument off, remove the batter
y
and wait for a minimum of one minute, reinsert the batter
y
and turn on. If the error message remains please contact the ThermoWorks’ Technical
Support for further assistance.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are
not possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item Non-contact Infrared Scan function
Thermocouple Probe Scan function
(T type; not included.)
Measurement Range
-67 to 536°F (-55 to 280°C)
-76 to 662°F (-60 to 350°C)
Operating Range
Accuracy, ( Amb=20 to 26°C)* -27.4 to 32°F (-33 to 0°C): ±1.8°F (±1°C) + 0.1°
32 to 122°F (0 to 50°C)
Accuracy
(Amb=20 to 26°C)
32 to 149°F (0 to 65°C): ±1.8°F (±1°C);
above 148°F (65°C): ±1.5%
Thermocouple Accuracy:
below 23°F (-5°C): ±1.8°F (±1°C),
23 to 148°F (-5 to 65°C): 0.9°F (±0.5°C),
above 148°F (65°C): ±1%
Resolution (-9.9 to 199.9°) 0.1°
Response Time (90%) 1 sec
Distance:Spot 11:1
Battery Life 180 hours typical 140 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions 6.9 H x 1.5 W x 2.8 D inches (175.2 H x 39.0 W x 71.9 D mm)
Weight 179 g including batteries (AAA*2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 18°F (10°C)
*IR accuracy can be affected by the emissivity setting, target spot size, and rapid changes in ambient temperature. Accuracy specification based on ambient
temperature of 77°F ±2°F (25°C ±1°C).
EMC/RFI
Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per
meter, but the performance of the instrument will not be permanently affected.
A new way to measure temperature
Radiant Innovation Inc. Http://www.ZyTemp.com
1F, No.3, Industrial East 9
th
Road, Science-Based Industrial
Park, HsinChu, Taiwan 300.
The IRT is a non-contact infrared thermometer for measuring surface temperature (infrared) and contact measurements
(optional probe). Please remember to keep away from
children and don’t use it for safety related applications.
(Default Screen)
9
MAIN FUNCTION
Press Emissivity key (
1
) for setting the emissivity.
Press Emissivity key (
1
), then press Up key (
4
) or Down key (
2
) to set the emissivity, then press Mode ke
y
(
5
) to confirm. The emissivity can be changed from 0.10 (10E) to 1 (100E) depending on the surface being measured.
Press Mode key (
5
) to scroll between the following functions:
Here will show the current emissivity value. The default emissivity is 0.95. For more information on emissivity
values of various surfaces visit www.thermoworks.com.
5
Press Up key ( ) or Down key ( ) key to change the High Alarm (HAL) or Lo Alarm (LAL) , then press
Meas. key ( ) to confirm. The alarm will sound if the readings are above the high alarm or below the low alarm.
4
2
9
Connect an optional thermocouple probe into the Thermocouple socket ( ) and insert the probe into the target
medium. The thermometer will display the temperature. To see the Min or Max temp during the probe
measurement, hold down the Up key ( ) or Down key ( ).
10
4
2
After measure high temp, the probe may remain HOT for a while.
** The thermometer will automatically shut off if left idle for more than 60sec, unless in PRB mode. (In PRB mode, it will shut off if
left idle for more than 12 minutes.)
ADDITIONAL FUNCTIONS
Changing from °C to °F Press the °C/°F down button
Display Lock Sets unit to continuous measure mode without having to press the trigger. Press Lock/Up button to
enable. Press again to disable.
Backlight While holding down the trigger (Meas. key), Press the Lock/Up button to turn backlight on. Press
again to disable.
Laser The laser pointer is set to on as the default. While holding down the trigger (Meas. key), press
the °C-°F/Down button to enable or disable the laser.
y
g
g
y
g
)
g
y
y
y
High Accuracy Infrared + Type T Thermocouple – Operating Instructions
I.R.T.
Use caution after measuring high temperatures, the probe may be hot.
T
T
N
N
4
4
0
0
8
8
X
X
C
C
/
/
T
T
N
N
4
4
0
0
8
8
L
L
C
C
T
T
h
h
e
e
r
r
m
m
o
o
m
m
e
e
t
t
e
e
r
r
O
O
p
p
e
e
r
r
a
a
t
t
i
i
n
n
g
g
I
I
n
n
s
s
t
t
r
r
u
u
c
c
t
t
i
i
o
o
n
n
s
s
Model: TN408XC, TN408LC (with Laser).
The thermometer is a non-contact infrared thermometer. There are many mathematical modes for the Infrared function. Please
remember to keep away from children and don’t use it
for safety related applications.
(Default Screen)
Simply aim the thermometer at the measure target
with Lens (
7
) and press Meas. key (
9
) to display
the surface temperature.
The Distance:Spot is 11:1. Please make sure the
target area is within the field of view.
FUNCTION
Press Emissivity key (
1
) for setting the emissivity.
Press Emissivity key (
1
), then press Up key (
4
) or Down key (
2
) to set the emissivity, then press Mode ke
y
(
5
) to confirm it. The emissivity can be changed from 0.10 (10E) to 1 (100E).
Press Mode key (
5
) for scrolling more display function as follows.
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key (
5
) for the Maximum (MAX), Minimum (MIN), Different between MAX and MIN (DIF) and
Average (AVG) modes. Durin
g
the measurement, the special modes readin
g
will be displa
y
ed beside the
mode icon.
Press Up key (
4
) or Down key (
2
) key to change the High Alarm (HAL) or Lo Alarm (LAL), then press Meas.
key (
9
) to confirm it. For example: When the reading 27
o
C < LAL 27.1
o
C, the Low icon will flash and
y
ou will hear
a beep sound.
Connect the thermocouple with Thermocouple socket (
10
) and put the probe in/on the tar
g
et, the thermometer will
display the temperature automatically without pressin
g
an
y
button. To see the minimum or maximum data durin
g
the probe measurement, please hold down the Up key (
4
) or Down key (
2
).
After measure high temp, the probe may remain HOT for a while.
** The thermometer will automatically shut off if left idle for more than 60sec, unless in PRB mode. (In PRB mode, it will shut off if
left idle for more than 12 minutes.)
ADD VALUE
Press Up key (
4
) for LOCK mode ON/OFF. The lock mode is particularl
y
useful for continuous
monitoring of temperatures for up to 60 minutes.
In MAX, MIN, DIF, AVG
mode:
Press Down key (
2
) for
o
C or
o
F transferred.
and press Up key (
4
) for backlight function ON/OFF.
In all modes: First
hold on the Meas. key
(
9
)
and press Down key (
2
) for laser function ON/OFF. (only for model TN408LC.)
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING The thermometer should be stored at room temperature between –20~+65
o
C (-4~149°F).
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the
thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lois displa
y
ed when the temperature bein
g
measured is outside of the settin
g
s of
HAL and LAL.
‘Er2’ is displa
y
ed when the thermometer is exposed to rapid chan
g
es in the ambient
temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0
o
C(32°F) or +50
o
C
(122°F). The thermometer should be allowed plenty of time (minimum 30 minutes
)
to stabilize
to the working/room temperature.
For all other error messa
g
es it is necessar
y
to reset the thermometer. To reset it, turn the
instrument off, remove the batter
y
and wait for a minimum of one minute, reinsert the batter
y
and turn on. If the error message remains please contact the Service Department for further
assistance.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are
not possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item Non-contact Infrared Scan function
Thermocouple Probe Scan function
(K type not included.)
Measurement Range
-60 to +500
o
C (-76 to +932
o
F)
-64 to +1400°C (-83.2 to +1999°F)
Operating Range 0~50°C (32~122°F)
Accuracy,
(Tobj=15-35°C,Tamb=25°C)
+/-1.0°C (1.8°F)
Accuracy
(Tobj=-33-500°C,Tamb=23 +/- 3°C)
+/-2% of reading or 2°C (4°F) whichever is greater
+/-1% of reading or 1°C (1.8°F)
whichever is
greater
(Test under Tamb=23+/-6°C)
Resolution
(-9.9~199.9°C) 0.1°C/0.1°F
Response Time
(90%) 1sec
Distance:Spot 11:1
Battery Life Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions 175.2 x 39.0 x 71.9mm
Weight 179 grams including batteries (AAA*2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10
o
C ( 18
o
F ).
EMC/RFI
Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per
meter, but the performance of the instrument will not be permanently affected.
Imported by
Made in Taiwan
1762 W. 20 S. #100, Lindon, UT 84042
Ph: 801-756-7705 Fax: 801-756-8948
www.thermoworks.com
Seitenansicht 0
1 2

Inhaltsverzeichnis

Seite 1

Press Mode key ( ) for the Maximum (MAX), Minimum (MIN), Difference between MAX and MIN (DIF)and Average (AVG) modes. During the measu

Seite 2 - CERTIFICATE OF CONFORMANCE

CERTIFICATE OF CONFORMANCEThe manufacturer of this instrument has implemented a quality assurance system under ISO 9001:2000 certified qualit

Kommentare zu diesen Handbüchern

Keine Kommentare